There are three settings common to every conventional mode of ventilation:
- FiO2:
- The amount of oxygen delivered to the patient
- PEEP:
- Pressure maintained in the respiratory system at the end of exhalation:
- Maintaining PEEP:
- Keeps an open lung and prevents atelectasis
- Maintaining PEEP:
- Pressure maintained in the respiratory system at the end of exhalation:
- Trigger sensitivity:
- Criteria used to see if the patient is making an effort:
- Flow-triggered
- Pressure-triggered
- Criteria used to see if the patient is making an effort:
- The three most common modes of ventilation include:
- Volume assist control
- Pressure assist control
- Pressure support
- Assist control modes (pressure or volume) are:
- Typically used in the acute phase of mechanical ventilation, or
- When the patient has no or very minimal drive to breath
- Pressure support:
- Is used when patients have an intact respiratory drive
- Volume Assist Control (AC-VC):
- Requires a frequency of respirations per minute
- Patients can trigger additional breaths greater than the devised respirations per minute
- If the trigger criteria is not being met, the machine will trigger all of the breaths
- When patients are starting to interact with the ventilator:
- A spontaneous mode:
- Such as pressure support should be considered
- A spontaneous mode:
- The tidal volume should be 6 to 8 mL/kg of ideal body weight:
- Current practice based on several trials suggests:
- That the patient should be ventilated with “lower” tidal volumes of 6 to 8 mL/kg
- Flow:
- Is the speed at which the tidal volume is delivered:
- 50 to 60 L / min will minimize discomfort when patients start making an effort
- Is the speed at which the tidal volume is delivered:
- PEEP should always be set at a minimum of:
- 5 cmH2O:
- To reduce atelectasis
- 5 cmH2O:
- The inspiratory flow:
- Is commonly set between 50 and 60 L/min and
- A minimum I:E ratio of:
- 1:1.5 to 1:2:
- Affected by respiratory rate as well
- 1:1.5 to 1:2:
- Common inspiratory times are:
- 0.75s to 1 s
- In certain circumstances:
- Such as in airway obstruction with asthma:
- Allowing more time for exhalation is beneficial:
- In these cases:
- One can increase the inspiratory flow or
- Decrease the I:E ratio:
- To 1:3 or 1:4
- In these cases:
- Allowing more time for exhalation is beneficial:
- Such as in airway obstruction with asthma:
- The inspiratory pause:
- Helps distinguish between:
- Resistive pressure and elastic pressure (compliance of the respiratory system)
- Allows the ventilator to display the:
- Plateau pressure:
- Which is helpful for monitoring the patient’s respiratory system mechanics (resistance and compliance)
- Plateau pressure:
- It also prolongs the inspiratory time to the common time of 0.75s to 1 s
- Helps distinguish between:
- Pressure Assist Control (AC-PC):
- Similarly requires a frequency of respirations per minute
- The inspiratory time:
- Is the length of time the pressure is maintained
- The rise time:
- Is the time the ventilator will take to reach the set pressure:
- The default setting for rise time is generally acceptable at:
- 0.1 sec
- The default setting for rise time is generally acceptable at:
- Is the time the ventilator will take to reach the set pressure:
- As resistance or elastance of the respiratory system changes:
- A result will be changes in:
- Tidal volume and minute ventilation
- Consequently, it is very important to monitor the tidal volume and keep it in the proper range in AC-PC
- Tidal volume and minute ventilation
- A result will be changes in:
- The I:E ratio:
- Is the simplest parameter to monitor if there is enough time to exhale:
- It must be maintained at 1:2 or higher to ensure there is enough time to exhale
- If the patient starts spontaneously breathing, this will affect I:E and it is worth considering transition to spontaneous breathing
- It must be maintained at 1:2 or higher to ensure there is enough time to exhale
- Is the simplest parameter to monitor if there is enough time to exhale:
- Similar to volume assist control:
- The inspiratory flow is commonly set at:
- 60 L/min
- Tidal volume should be:
- 6 to 8 mL/kg
- The inspiratory flow is commonly set at:
- Pressure Support (PS):
- Is distinguished from AC-PC:
- Because breaths are cycled off by a % of peak flow:
- As opposed to time
- Because breaths are cycled off by a % of peak flow:
- It is important to adjust the default settings for the % of peak flow to initiate the cycling off of each breath:
- As it depends highly on the resistance and elastance of the lungs
- Is distinguished from AC-PC:
Levels to monitor:
- Volume Assist Control:
- Elevated peak pressure and/or plateau pressure:
- Will result from abnormal resistance and elastance due to:
- ARDS, COPD, asthma, intra-abdominal hypertension, etc.
- Will result from abnormal resistance and elastance due to:
- Elevated peak pressure and/or plateau pressure:
- Pressure Assist Control:
- Significant changes to tidal volume and minute ventilation:
- Will result from abnormal resistance and elastance due to:
- ARDS, COPD, asthma, intra-abdominal hypertension, etc.
- Will result from abnormal resistance and elastance due to:
- Significant changes to tidal volume and minute ventilation:
- PEEP:
- Can be increased to improve oxygenation:
- However, it should not be so much to overdistend the lungs
- The risk of potential injury increases:
- When plateau pressure is greater than:
- 27 cm H2O
- When plateau pressure is greater than:
- Can be increased to improve oxygenation:
- pH should be kept between 7.35 and 7.45:
- To increase pH:
- Increase minute ventilation
- To decrease pH:
- Decrease minute ventilation
- However:
- Minute Ventilation should not be increased to the point that PaCO2 < 30 mmHg:
- Cerebral perfusion may be impacted with levels that low
- Minute Ventilation should not be increased to the point that PaCO2 < 30 mmHg:
- In management of ARDS:
- Permissive hypercapnia is often considered to minimize injury to the lung and a pH 7.25 is considered the lower acceptable limit
- To increase pH:
#Arrangoiz #Surgeon #Teacher
